Page 80 QUICK300 CONTROL SOFTWARE
C
YBER
A
MP
380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC.
port speed should be set to fast unless a communications problem is experienced. To accept the
changes in this or any dialog form, exit the form with the keystroke <Enter>. To reject the changes,
exit the form with <Esc>. The COM port number and the communication speed are saved in the file
QUICK300.INI.
The first time that QUICK300 is run, it will set the CyberAmp to its factory default settings. When
the program is terminated with eXit, it will save the current CyberAmp settings in a file named
QUICK300.INI. The next time the program is run, it will restore the CyberAmp to the values in this
file. Note that settings of the CyberAmp made in other programs, such as CMD300, will not be
included in this file and will be lost when QUICK300 is next run. A DOS command line option
permits QUICK300 to be loaded without overwriting the existing CyberAmp settings (the /l option).
See Command Line Options below.
Return to the main menu with the keystroke <Enter>. Choose SetValues to display the control
spreadsheet.
S=(SETUP1) Connection enabled
Program QUICK300 Version 2.0
Copyright (c) Axon Instruments Inc. 1984-1992. All rights reserved.
-- CyberAmp Setup --------------------------------------------------
CyberAmp chan: #1 #2 #3 #4 #5 #6 #7 #8
Probe model: AI417 AI401 AI402 AI418 AI405 AI410 AI490 AI417
Positive input: DC DC DC DC GND DC AC AC
Negative input: DC GND GND GND DC DC AC AC
AC cutoff (Hz): GND GND GND GND GND GND 10 0.1
Pre-filter gain: 100 100 100 1 20 10 10 1
Offset (mV): 0.80 -1.40 0.20 0 2.60 0 0 0
Lowpass (Hz): 600 2800 6000 1 Bypass 100 Bypass 1000
Notch filter: OFF OFF OFF OFF OFF OFF OFF ON
Output gain: 50 200 100 1 10 100 100 50
Total gain: 5000 20000 10000 1 200 1000 1000 50
Use <F5> and <Shift+F5> to copy rows and columns, <PgUp> and <PgDn> to
select.
Enter the CyberAmp gain
Each channel is represented by a column and each function is represented by a row. Using the
computer keyboard, the arrow keys, <Tab> and <Shift-Tab> move the highlight to the desired cell of
the spreadsheet. The bottom line provides help for each cell. The top line includes the name of a set-
up file, if one has been loaded.
The selected function is changed by scrolling or by keyboard entry. The <PgUp> and <PgDn> keys
are used to scroll through permissible values of the function. The change in the selected CyberAmp
function is immediate. Alternatively, the value can be entered from the keyboard, followed by
<Enter>. If the entered value is permissible, the CyberAmp will be changed; otherwise, the previous
value is restored.