CMD300 CONTROL SOFTWARE Page 73
C
YBER
A
MP
380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC.
12)
O
FFSET <offset value>
The OFFSET command specifies the DC offset adjustment that is added after the pre-filter
gain amplifier. The <offset value> is specified in millivolts and can be in the range -3000.0 to
+3000.0 mV. Because this offset is introduced after the pre-filter gain amplifier, the input-
referred offset value depends on the gain setting of the pre-filter amplifier.
13)
OV
ERLOAD
OVERLOAD commands the CyberAmp to report which channels have overloaded since the
last time that this command was issued or since the unit was powered up. After the report, the
CyberAmps resets the overload status for all channels. The following response indicates that
channels 1, 3, 4 and 7 have overloaded since the last check.
CMD300 OVERLOAD
CyberAmp overload status: 1, -, 3, 4, -, -, 7, -
14)
PO
STGAIN <output gain value>
This command sets the output gain amplifier for the selected channels without changing the
pre-filter gain amplifier. Allowable values for <output gain value> are:
1, 2, 5, 10, 20, 50, 100 and 200.
15)
PR
EGAIN <pre-filter gain value>
This command sets the pre-filter gain amplifier for the selected channels without changing the
output gain amplifier. Allowable values for <pre-filter gain value> are:
1, 10 and 100.
16)
QUIET
This command minimizes all subsequent CyberAmp responses to commands. In QUIET
mode no confirmation is given for simple commands such as changes to gain or filter settings.
CyberAmp responses are maximized with the VERBOSE Command.
17)
STATUS
or
STATUS DEFAULTS
or
STATUS ID
The STATUS command displays the CyberAmp unit type, serial number, firmware revision
number, and status information for all channels. The format of the channel status information
is as follows:
#1 Pr=AI401 +=DC -=Gnd Gain=20(10,2) Filt=10k Notch=N Offset=+50mV