Page 102 PROGRAMMER'S GUIDE
C
YBER
A
MP
380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC.
"L" - Load factory defaults
Format
ATdL
where
d = device # (0-9)
Notes
All channels are simultaneously set to the factory defaults.
The factory defaults are:
+ input = DC
- input = Ground
Pre-filter gain = x1
Output gain = x1
Low-pass filter = 10 kHz
Offset = zero
Notch filter = out
"N" Notch filter
Format
ATdNnp
where
d = device # (0-9)
n = channel number (1-8)
p = "+" if the notch filter is included in the signal pathway
"-" if the notch filter is removed from the signal pathway
Examples
AT3N1+
Add notch filter to device 3, channel #1
AT3N1-
Remove notch filter