Page 12 FUNCTIONAL CHECKOUT
CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC.
The computer will respond with the following status information:
Program CMD300 Version 1.nnn
Copyright (c) Axon Instruments Inc. 1990. All rights reserved.
CyberAmp reset to factory defaults.
CYBERAMP 380 (Serial number nnnn) (Microcode version 1.nnn)
#1 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0V
#2 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0V
#3 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0V
#4 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0V
#5 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0V
#6 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0V
#7 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0V
#8 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0V
These are the factory default settings.
6) Connect the vertical input of your oscilloscope to the Channel 1 output. Set the oscilloscope
controls to 0.2 V/div, 20 ms/div. Issue the following command to the CyberAmp 380 to
generate a test waveform on all channels:
CMD300 TEST ELECTRODE ON
The oscilloscope should display a 1 V
p-p
square wave at 10 Hz (50 ms high, 50 ms low).
Move the BNC cable to channels 2 through 8. The same waveform should be on each
channel.
7) Issue the command:
CMD300 CHANNEL ALL FILTER 6
to set the low-pass filter on all channels to 6 Hertz. The waveform should look almost
sinusoidal and about 0.4 V
p-p
. Verify this for all channels.
8) Issue the command:
CMD300 GAIN 10
to set the gain to x10. The waveform amplitude on each channel should increase ten-fold to
about 4 V
p-p
.